calibration problem


Switching the marker when you have calibration problems

Back to the list of tips



When calibrating the field deflection with "aeb" or

pg adjust ebpg

we sometimes find that pull-in compensation exceeds the limit, even after several iterations. The error message looks like this:

defl = 522.060000,522.060000
pi (x,y) = 3.481_um, 0.000_um err = -26.8_nm, 16.8_nm
pi (x,y) = -3.481_um, -0.000_um err = -25.4_nm, 11.8_nm
pi (x,y) = -0.000_um, 3.481_um err = -26.6_nm, 8.2_nm
pi (x,y) = 0.000_um, -3.481_um err = -27.3_nm, 11.6_nm
PICOM = 2.0_nm, measured = 3.6_nm
 
%ENG_E_NOOPRE - the optima for the fine focus,
        diagonal and axis stigmators are not reproducible

This error might be caused by a dirty marker. You can switch to a new marker, temporarily, by centering a new mark in SEM mode, then using the command

usethismark

This command is a little script that runs these two commands:

mvm 0,0 --rel p20
pg marker calib `pg get tab` p20

After switching marks, try "aeb" again. If this fixes the problem, then we should update the marker coordinate in the holder table, so that the new mark will be used by the next users. You should ask the system administrator to update the holder table.